Recovery of the Hip Rotation Center with Tantalum in Revision Arthroplasty*

Recuperação do centro de rotação do quadril com tântalo em artroplastias de revisão*

Antônio Augusto Guimarães Barros1,, Victor Atsushi Kasuya Barbosa1, Lincoln Paiva Costa1, Euler de Carvalho Guedes1, Carlos César Vassalo1

DOI: 10.1055/s-0039-1693047


 

ABSTRACT:

OBJECTIVE The objective of the present study is to evaluate the restoration capacity of the hip anatomic rotation center with the use of acetabular tantalum cups, associated or not with addition wedges.
METHODS Retrospective analysis of patients undergoing hip arthroplasty revision using tantalum between June 2013 and April 2017. The abduction angle of the acetabular component and the horizontal and vertical distances of the component to the center of anatomical rotation of the hip were evaluated. The measurements were made through baseline radiographs performed in the preoperative period and at the last follow-up visit.
RESULTS A sample of 21 patients was obtained, 11 (52%) men and 10 (48%) women, with a mean age of 62 ± 13 years old. The mean abduction angle of the acetabular cup decreased from 48.76º ± 13.88 º in the preoperative period to 38.52º ± 10.08 º in the postoperative period, and this difference was statistically significant (p = 0.001). The distances from the center of rotation of the prosthesis relative to the center of anatomical rotation of the hip were also lower after revision surgery with tantalum. The mean horizontal distance of 12.74 ± 10.59 mm was reduced to 7.11 ± 4.84 mm, and the mean vertical distance was reduced from 14.79 ± 10.05 mm to 4.89 ± 6.21 mm, and these reductions were statistically significant (p < 0.001).
CONCLUSION Hip arthroplasty revision with tantalum cups, associated or not with addition wedges, significantly recovered the anatomical rotation center of the hip.

Keywords:
arthroplasty, replacement, hip; acetabulum; tantalum.

RESUMO:

OBJETIVO O objetivo do presente estudo é avaliar a capacidade de restauração do centro de rotação anatômico do quadril com uso de copas acetabulares de tântalo associado ou não a cunhas de adição.
MÉTODOS Análise retrospectiva dos pacientes submetidos a revisão de artroplastia do quadril com uso de tântalo entre o período de junho de 2013 e abril de 2017. Foram avaliados o ângulo de abdução do componente acetabular e as distâncias horizontal e vertical do componente ao centro de rotação anatômico do quadril. As medidas foram realizadas através de radiografias da bacia realizadas no pré-operatório e na última visita de seguimento.
RESULTADOS Obteve-se uma amostra de 21 pacientes, 11 (52%) homens e 10 (48%) mulheres, com média de idade de 62 ± 13 anos. O ângulo médio de abdução da copa acetabular reduziu de 48,76º ± 13,88º no pré-operatório para 38,52º ± 10,08º no pós-operatório, sendo esta diferença estatisticamente significativa (p = 0,001). As distâncias do centro de rotação da prótese em relação ao centro de rotação anatômico do quadril também foram menores após a cirurgia de revisão com o tântalo. A distância média horizontal de 12,74 ± 10,59 mm foi reduzida para 7,11 ± 4,84 mm, e a distância média vertical foi reduzida de 14,79 ± 10,05 mm para 4,89 ± 6,21 mm, sendo essas reduções estatisticamente significativas (p < 0,001).
CONCLUSÃO As revisões de artroplastia do quadril com copas de tântalo, associadas ou não a cunhas de adição, recuperaram de forma significativa o centro de rotação anatômico do quadril.

Palavras-chave:
artroplastia de quadril; acetábulo; tântalo.

FIGURES

Citation: Barros AAG, Barbosa VAK, Costa LP, Guedes EC, Vassalo CC. Recovery of the Hip Rotation Center with Tantalum in Revision Arthroplasty*. 54(4):471. doi:10.1055/s-0039-1693047
Note: * Work performed at the Hospital Madre Teresa, Belo Horizonte, MG, Brazil.
Received: March 14 2018; Accepted: August 21 2018
 

INTRODUCTION

The number of revision total hip arthroplasty (rTHA) procedures tends to increase in the coming years. In the USA, the number of rTHAs is expected to double between 2005 and 2026.1 The most common causes for revision are instability (17.3%), aseptic release (16.8%), unspecified mechanical complications (13.4%), and infection (12.8%).2 In most cases, the acetabular component is replaced, whereas all components are changed in 40.3% of the cases, and the acetabulum alone is replaced in 14.5% of reoperations.2 The revision of this component with restoration of the anatomical hip rotation center, acetabular bone defects correction, and primary crown stability is a challenge for the orthopedic surgeon.3 The most commonly described techniques are the use of large or extra-large conventional crowns without cement,4 cemented cups with or without impacted graft,5 acetabular reinforcing ring,6 cup-cage construction,7 and the use of cups with high-porosity metals (e.g., tantalum) associated or not with addition wedges.8 Despite the amount of available options, these techniques still present variable results according to the degree of bone lesion, with a failure rate of up to 88.5% of cases after a mean follow-up of 44.6 months.9

Even in cases of complex revisions with great bone loss, revision implants coated with high-porosity metal have good results in short- and medium-term follow-ups.10,11 This material presents theoretical advantages compared with other materials, such as high friction coefficient, modulus of elasticity similar to bone, and a large porosity volume that allows greater osseointegration and better secondary crown fixation.12,13 However, these are relatively new, expensive materials, and a considerable burden to the health system. The present study aims to evaluate the restoration capacity of the anatomical hip rotation center using acetabular crowns coated with high-porosity metal, associated or not with addition wedges. Our hypothesis is that rTHAs performed with high-porosity metal are effective in improving the positioning of the hip rotation center when compared with the values recommended by the literature.

 

MATERIALS AND METHODS

Using the arthroplasty database from the hospital of our institution, all of the patients who underwent rTHAs between June 2013 and April 2017 were identified. After approval by the institutional ethics committee, patients submitted to acetabular surgical reconstruction using tantalum-coated components, associated or not to addition wedges (Trabecular Metal, Zimmer, Warsaw, Indiana, USA), were invited to participate in a retrospective analysis. The present study included all patients who underwent THAs using tantalum at the hospital between June 2013 and April 2017, regardless of the reason for the revision, of the degree of bone loss, or of the presence of any comorbidity. Only patients who died or did not perform an adequate segment were excluded.

Defects prior to the surgery were described according to the classification of Paprosky et al,14 and the radiographic parameters were measured by hip radiographs (anteroposterior [AP] and oblique views) performed during the preoperative periods and at the last follow-up visit. The abduction angle of the acetabular component, in relation to the teardrop or ischial tuberosity, as well as the horizontal and vertical distances between the component and the anatomical hip rotation center, were measured. From the pelvic height, an isosceles triangle was outlined at 5 mm laterally from the intersection between the Shenton line and the Koehler line, with its sides measuring 20% of the pelvic height. The rotation center is defined as half the length of the hypotenuse.15

 

STATISTICAL ANALYSIS

Categorical variables were presented as absolute and percentual values, whereas continuous variables were shown as mean and standard deviation (SD). Continuous variables were submitted to a normality evaluation by the Shapiro-Wilk test. The comparison between the pre- and postoperative periods was performed with the two-tailed Student t-test for variables in parametric distribution. For independent samples, the comparison between the two groups was performed through a proper Student t-test after confirming the normal distribution and evaluating variances by the Levene test. The correlation coefficient between pre- and postoperative values for continuous variables was determined by the Spearman test. The correlation power was classified according to the r value, being strong if > 0.70, moderate between 0.30 and 0.70, and weak between 0 and 0.30. Data was analyzed with SPSS Statistics for Windows, Version 20.0 (IBM Corp., Armonk, NY, USA), with a 5% significance level.

 

RESULTS

Between June 2013 and April 2017, 95 revisions of hip arthroplasty were performed. Of this total, 21 (22%) patients underwent acetabular revision with a tantalum-coated crown, associated or not to addition wedges (Zimmer, Warsaw, IN, USA). A trend for increasing use of this material was observed in our sample (Fig. 1). The characteristics of the sample are shown in Table 1.

Table 1. Sample characteristics
Patients, n 21
Gender, n (%)
Male 11 (52.4%)
Female 10 (47.6%)
Mean age, years old (standard deviation) 62 (±13.2)
Revision indication, n (%)
Aseptic release 15 (71.4%)
Instability 1 (4.8%)
Infection 3 (14.3%)
Pain 2 (9.5%)
Number of previous revisions, n (%)
0 11 (52.4%)
1 8 (38.1%)
2 2 (9.5%)
Paprosky classification, n (%)
Type 1 3 (14.3%)
Type 2ª 3 (14.3%)
Type 2B 6 (28.6%)
Type 2C 0 (0%)
Type 3ª 2 (9.5%)
Type 3B 2 (9.5%)
Pelvic discontinuity 5 (23.8%)

The mean abduction angle of the acetabular cup decreased from 48.76 ± 13.88º at the preoperative period to 38.52 ± 10.08º at the postoperative period, and this difference was statistically significant (p = 0.001). The distances between the rotation center of the prosthesis and the anatomical hip rotation center were also shorter after the revision surgery using an acetabular tantalum component. The mean horizontal distance of 12.74 ± 10.59 mm was reduced to 7.11 ± 4.84 mm, and the mean vertical distance was reduced from 14.79 ± 10.05 mm to 4.89 ± 6.21 mm; both reductions were statistically significant (p < 0.001) (Table 2).

Table 2. Radiological results of hip mechanics
  Preoperative Postoperative p-value
Acetabular cup abduction angle*, mean ± standard deviation 48.76 ± 13.88 38.52 ± 10.08 p = 0.001a
Horizontal distance**, mean ± standard deviation 12.74 ± 10.59 7.11 ± 4.84 p < 0.001a
Vertical distance**, mean ± standard deviation 14.79 ± 10.05 4.89 ± 6.21 p < 0.001a

* Value in degrees.

** Value in mm.

a Two-tailed Student t-test.

There was a strong direct correlation between the pre- and postoperative values of horizontal distance (r = 0.928; p < 0.001) and of vertical distance (0.792; p < 0.001). Therefore, the presence of a greater distance to the anatomical hip rotation center before the revision was also associated to a greater distance after the surgery, indicating that major rotation center deviations present smaller postoperative corrections (Table 3).

Table 3. Correlation between preoperative and postoperative radiological valuesa
  R value p-value
Acetabular cup abduction angle 0.525 p = 0.015
Horizontal distance 0.928 p < 0.001
Vertical distance 0.792 p < 0.001

a Spearman correlation coefficient.

Comparing the postoperative radiographic results between both groups without previous revision (11 patients) and with previous revision (10 patients), no statistically significant difference was observed in the crown abduction angles or in the distances to the anatomical hip rotation center (Table 4).

Table 4. Comparison of postoperative radiological results between the groups with and without previous revisions
  Absence of previous revision Presence of previous revision p-value
Acetabular cup abduction angle*, mean ± standard deviation 36.09 ± 11.29 41.20 ± 8.29 p = 0.256a
Horizontal distance**, mean ± standard deviation 6.36 ± 3.66 7.94 ± 5.98 p = 0.470a
Vertical distance**, mean ± standard deviation 3.65 ± 6.37 6.24 ± 6.06 p = 0.354a

* Value in degrees.

** Value in mm.

a Student t-test for independent samples.

 

DISCUSSION

The present study showed that the use of tantalum-coated acetabular cups, associated or not with addition wedges, was effective in recovering the anatomical hip rotation center in rTHA surgeries (Figs. 2a, 2b, 3a and 3b). Thus, the hypothesis of the present study was corroborated.

Correct acetabular positioning is fundamental to the success of total hip arthroplasty (THA), which may influence the load inflicted to the joint and implant wear. Regarding the anatomical hip rotation center, each millimeter of crown lateralization and proximalization is associated with an increase of 0.7% and of 0.1% in the joint load, respectively.16 There is some consensus in the literature that the acetabular crown should be up to a maximum distance of 5 mm from the anatomical hip rotation center.17-19 Cup positioning with an abduction angle > 45º was associated with a 40% increase in the average linear wear of polyethylene.20 A more horizontal acetabular cup positioning may reduce pelvic osteolysis. Kennedy et al21 assessed 75 patients submitted to THA who were divided in 2 groups with mean abduction values of 61.9º and 49.7º, respectively; these authors observed pelvic osteolysis rates of 24% and of 13% in higher and lower acetabular inclination groups, respectively, after 4 years of follow-up.

The improvement of the position of the hip rotation center may be beneficial to the functional results,22 since the center of rotation is very important for muscle function. Asayama et al,23 when studying 30 patients submitted to THA, found a negative correlation between abductor musculature strength and the proximalization of the hip rotation center. Another study of biomechanical analysis in a lower limb model showed that a 2 cm superior deviation was related to a 44% decrease in abductor strength.24 The weakness of the abductor hip muscle was one of the most important causes of prosthesis dislocation in a study involving 1,318 patients.25

Abolghasemian et al26 corroborated the results of the improvement of the positioning of the acetabular component rotation center after revision with a high-porosity metal. The position of the rotation center improved in both axes after the revision; the mean distance to the anatomical center prior to the revision was 28.8 mm (- 3 to 79 mm) on the vertical axis, and 13.3 mm (- 21 to 35 mm) on the horizontal axis. At the postoperative period, the mean distance from the rotation center was 9.9 mm (- 18 to 37 mm) superior, and 5.1 mm (- 25 to 30 mm) lateral to the anatomical center. Jenkins et al10 also reported recovery of the hip rotation center. The mean horizontal distance to the hip rotation center decreased from 9 mm preoperatively (range: 1 to 45 mm) to 8 mm postoperatively (range: 0 to 27 mm) (p = 0.0143), and it was 9 mm (range: 0 to 26 mm) at the last follow-up visit. The mean vertical distance to the hip rotation center decreased from 21 mm preoperatively (range: 0 to 80 mm) to 6 mm postoperatively (range: 0 to 28 mm) (p = 0.0001), and it was 4.5 mm (range: 0 to 57 mm) at the last follow-up visit. The mean abduction angle of the acetabular cup was 61º at the preoperative period (range: - 18º to 180º), 45º postoperatively (range: 34º to 60º) (p = 0.0001), and 45º (range: 34º to 82º) at the last follow-up visit.

The present study has some limitations. It is a retrospective study with a relatively small sample of 21 patients. The restoration of the rotation center was evaluated only in the coronal plane, not considering anteroposterior deviations; moreover, the acetabular version was not evaluated.

 

CONCLUSION

Tantalum-coated acetabular cups, associated or not with addition wedges, were effective in significantly improving the positioning of the anatomical hip rotation center in revision surgeries.

 

REFERENCES

Kurtz S, Ong K, Lau E, Mowat F, HalpernM. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 2007;89(04):780–785 Link DOI
GwamCU,Mistry JB,Mohamed NS, ThomasM, Bigart KC,Mont MA, et al. Current Epidemiology of Revision Total Hip Arthroplasty in the UnitedStates:National Inpatient Sample2009to2013. JArthroplasty 2017;32(07):2088–2092 Link DOI
Paprosky WG, Magnus RE. Principles of bone grafting in revision total hip arthroplasty. Acetabular technique. Clin Orthop Relat Res 1994;(298):147–155 Link DOI
Lachiewicz PF, Soileau ES. Fixation, survival, and dislocation of jumbo acetabular components in revision hip arthroplasty. J Bone Joint Surg Am 2013;95(06):543–548 Link DOI
Schreurs BW, Keurentjes JC, Gardeniers JW, Verdonschot N, Slooff TJ, Veth RP. Acetabular revision with impacted morsellised cancellous bone graftinganda cemented acetabular component: a 20- to25-year follow-up. J Bone Joint Surg Br 2009;91(09):1148–1153 Link DOI
Mäkinen TJ, Kuzyk P, Safir OA, Backstein D, Gross AE. Role of cages in revision arthroplasty of the acetabulum. J Bone Joint Surg Am 2016;98(03):233–242 Link DOI
Mäkinen TJ, Fichman SG,Watts E, Kuzyk PR, Safir OA, Gross AE. The role ofcages inthemanagementof severe acetabularbonedefects at revision arthroplasty. Bone Joint J 2016;98-B(1, Suppl A):73–77 Link DOI
Laaksonen I, Lorimer M, Gromov K, Rolfson O,Mäkelä KT, Graves SE, et al. Does the riskof rerevisionvary between porous tantalumcups and other cementless designs after revision hip arthroplasty? Clin Orthop Relat Res 2017;475(12):3015–3022 Link DOI
Kosashvili Y, Backstein D, Safir O, Lakstein D, Gross AE. Acetabular revision using an anti-protrusion (ilio-ischial) cage and trabecular metal acetabular component for severe acetabular bone loss associated with pelvic discontinuity. J Bone Joint Surg Br 2009; 91(07):870–876 Link DOI
Jenkins DR, Odland AN, Sierra RJ, Hanssen AD, Lewallen DG. Minimumfive- year outcomes with porous tantalumacetabular cup and augment construct in complex revision total hip arthroplasty. J Bone Joint Surg Am 2017;99(10):e49 Link DOI
Konan S, Duncan CP, Masri BA, Garbuz DS. Porous tantalum uncemented acetabular components in revision total hip arthroplasty: a minimum ten-year clinical, radiological and quality of life outcome study. Bone Joint J 2016;98-B(06):767–771 Link DOI
Bobyn JD, Stackpool GJ, Hacking SA, Tanzer M, Krygier JJ. Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial. J Bone Joint Surg Br 1999;81(05):907–914 Link DOI
Bobyn JD, Poggie RA, Krygier JJ, Lewallen DG, Hanssen AD, Lewis RJ, et al. Clinical validation of a structural porous tantalumbiomaterial for adult reconstruction. J Bone Joint Surg Am 2004;86-A (Suppl 2):123–129 Link DOI
Paprosky WG, Perona PG, Lawrence JM. Acetabular defect classification and surgical reconstruction in revision arthroplasty.A6-year follow-up evaluation. J Arthroplasty 1994;9(01):33–44 Link DOI
Pagnano W, Hanssen AD, Lewallen DG, Shaughnessy WJ. The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty. J Bone Joint Surg Am 1996;78(07):1004–1014
Bicanic G, Delimar D, Delimar M, Pecina M. Influence of the acetabular cup position on hip load during arthroplasty in hip dysplasia. Int Orthop 2009;33(02):397–402 Link DOI
Liebs TR, Nasser L, Herzberg W, Rüther W, Hassenpflug J. The influence of femoral offset on health-related quality of life after total hip replacement. Bone Joint J 2014;96-B(01):36–42 Link PubMed Link DOI
Cassidy KA, Noticewala MS, Macaulay W, Lee JH, Geller JA. Effect of femoral offset on pain and function after total hip arthroplasty. J Arthroplasty 2012;27(10):1863–1869 Link DOI
Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 2002;17(03):282–288 Link DOI
Patil S, Bergula A, Chen PC, Colwell CWJr,D’Lima DD. Polyethylene wear and acetabular component orientation. J Bone Joint Surg Am 2003;85-A(04, Suppl 4):56–63 Link DOI
Kennedy JG, RogersWB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty 1998;13(05):530–534 Link DOI
KimDH, Cho SH, Jeong ST, Park HB,Hwang SC, Park JS. Restoration of the centerof rotationinrevisiontotalhiparthroplasty. JArthroplasty 2010;25(07):1041–1046 Link DOI
Asayama I, Chamnongkich S, Simpson KJ, Kinsey TL, Mahoney OM. Reconstructed hip joint position and abductor muscle strength after total hip arthroplasty. J Arthroplasty 2005;20 (04):414–420 Link DOI
Delp SL, Maloney W. Effects of hip center location on the momentgeneratingcapacityof themuscles. JBiomech1993;26(4-5):485–499 Link DOI
García-Rey E, García-Cimbrelo E. Abductor biomechanics clinically impact the total hip arthroplasty dislocation rate: a prospective long-term study. J Arthroplasty 2016;31(02):484–490 Link DOI
Abolghasemian M, Tangsataporn S, Sternheim A, Backstein D, Safir O, Gross AE. Combined trabecular metal acetabular shell and augment for acetabular revision with substantial bone loss: a mid-term review. Bone Joint J 2013;95-B(02):166–172 Link DOI

*Trabalho realizado no Hospital Madre Teresa, Belo Horizonte, MG, Brasil.