Original Article

FIXATION OF SUPRACONDYLAR FEMORAL FRACTURES: A BIOMECHANICAL ANALYSIS COMPARING 95O BLADE PLATES AND DYNAMIC CONDYLAR SCREWS (DCS)

AVALIAÇÃO BIOMECÂNICA DA FIXAÇÃO DA FRATURA SUPRACONDILIANA DO FÊMUR COMPARANDO PLACA-LÂMINA 95O COM DCS

Marco Antônio Percope Andrade1, André Soares Rodrigues2, Celso Junio Mendonça3, Luiz Gustavo Santos Portela3



ABSTRACT

ABSTRACT Objective: To determine, by means of comparative biomechanical tests, whether greater compressive load resistance and flexion is presented by 95o angled blade plates or by dynamic condylar screws (DCS), and to correlate the failure type presented during the tests with each type of plate. Methods: Sixty-five porcine femurs were subjected to 1 cm medial wedge osteotomy, in the metaphysis, to simulate an unstable supracondylar femoral fracture. Osteosynthesis was performed on these pieces: 35 were fixed using 95o lateral blade plates and 30 with DCS plates. Another variable studied was the failure type presented in each group, in an attempt to correlate this with the type of plate. Results: There were no statistically significant differences in biomechanical resistance between the two types of plates, or between the failure type and the plate type used for the osteosynthesis. Conclusion: The two types of plate behaved in a similar fashion. However, the angled blade plate proved to be superior to the DCS in the flexion test. No statistical difference in failure type or type of plate used was observed. Keywords - Femoral fractures; Fracture fixation, internal; Biomechanics

RESUMO

Objetivo: Demonstrar por meio de ensaios biomecânicos comparativos entre a placa-lâmina de 95º e o parafuso condilar dinâmico (Dynamic Condylar Screw - DCS), qual apresenta maior resistência às cargas compressivas e de flexão, bem como tentar correlacionar o tipo de falha apresentada durante os testes com cada um dos tipos de placa. Métodos: Sessenta e cinco fêmures suínos foram submetidos a osteotomia em cunha de subtração medial de um centímetro (cm), na região metafisária distal do fêmur, com o objetivo de simular fratura supracondiliana instável. Foi realizada osteossíntese dessas peças, sendo 35 fixadas com placa-lâmina 95º e 30 com placas com DCS, submetendo-as a cargas em compressão axial e flexão. Outra variável estudada foi o tipo de falha apresentada em cada grupo com a tentativa de correlacioná-la com o tipo de placa. Resultados: Os resultados não mostraram diferença estatisticamente significante na resistência biomecânica entre os dois tipos de placas ou entre o tipo de falha e a placa utilizada na osteossíntese. Conclusão: Os dois tipos de placas se comportam de maneira semelhante, embora haja um indicativo de que a placa-lâmina seja, no ensaio de flexão, superior à placa DCS. Não foi observada diferença entre o tipo de falha e o tipo de placa utilizada. Descritores - Fraturas do fêmur; Fixação interna de fraturas; Biomecânica.

Untitled Document

1 - Head of the Orthopedics and Traumatology Service of Hospital das Clínicas, Federal University of Minas Gerais (UFMG). Adjunct Professor in the School of Medicine, UFMG.
2 - Orthopedist and Specialist in Hip Surgery. Member of the Clinical Staff of Risoleta Tolentino Neves University Hospital, UFMG.
3 - Former Resident at Hospital das Clínicas, Federal University of Minas Gerais (UFMG).
Work developed at Hospital das Clínicas and on the campus of the Federal University of Minas Gerais.
Correspondence: André Soares Rodrigues - Rua Antonio de Albuquerque, 877, apto 1.102, Funcionários - 30112-011 - Belo Horizonte, MG - E-mail: andresoaresrodrigues@hotmail.com

INTRODUCTION

Fractures of the supracondylar region of the femur are complex lesions that are usually difficult to treat. They correspond to approximately 7% of all fracture of the femur and to 31% of them, if fractures of the proximal femur are excluded(1). The most frequent cause among the elderly population is falls from standing height with the knee flexed, while among young patients, it is high-energy trauma (traffic accidents and falls from a height), which generally lead to varus, valgus or rotational forces with axial loading. Ligament injuries may be associated with around 20% of these cases(1), along with fractures of the acetabulum, the femoral neck or diaphysis and the tibial plateau.

One peculiar characteristic of such fractures is the deformation caused by the different muscle groups that act on the knee (quadriceps, ischiotibial, gastrocnemius and adductor muscles). This leads to deviation of the fragments, particularly in situations of hyperextension due to the action of the gastrocnemius, with the need for open reduction and internal fixation.

Controversy still exists concerning which method is best for supracondylar fractures of the femur. Several implantation methods are used for fixation of this type of fracture, but without any consensus regarding the method that would be most stable biomechanically. In addition to retrograde intramedullary fixation, plates with blades angled at 95º and dynamic condylar screws (DCS) have been highlighted.

Blade plates provide excellent fixation and are considered to be the method presenting greatest resistance to angulation and torsion forces, despite the greater technical difficulty(2-3). On the other hand, because DCS are thicker than blade plates, they theoretically cause greater bone destruction(2). Studies comparing DCS with retrograde and anterograde intramedullary rods have shown conflicting results(4-7).

The aim of this study was to compare the biomechanical rigidity of internal fixation performed using 95º blade plates and DCS, in a simulation of unstable extra-articular supracondylar fractures in porcine femurs, and to correlate the types of plate with the type of failure, i.e. whether this was in the bone (fracture) or in the material used (loosening on breakage of the implant). These implants were chosen because they were the types most commonly used for treating the type of fracture that we study in our setting.

METHODS

Sixty-five femurs from pigs of the Landraz breed, of mean age 90 days, were selected for the experiment. The bones were stored at a temperature of -18ºC and then were placed at room temperature 12 hours before the fixation procedure.

Sixty-five plates were used for fixation of the femurs. Al of them had five orifices: 35 with a blade angled at 95º, of 55 millimeters (mm) in length, and 30 with DCS of 55 mm in length. The fixation at the most distal hole in both plates was performed using a spongy bone screw and three cortical bone screws were used in the proximal holes, while the hole corresponding to the region of the osteotomy was left without fixation. The osteotomy was performed around 4 cm from the joint surface. For osteosynthesis, the principles and techniques of the AO/ ASIF group were used(8), using plates and screws produced by the company I.O.L. Implantes Ltda.

After fixation of the femurs, wedge osteotomy was performed, with medial subtraction of 1 cm of material in the distal metaphysis region of the femur, using a oscillatory nitrogen saw, with the aim of creating an unstable supracondylar fracture (without medial support) (Figure 1).

The bones were again stored at a temperature of approximately -18ºC after fixation. Twenty hours before starting the experiment, they were transferred to a refrigerator at a temperature of around 4ºC, and then, one hour before the tests, they were left at room temperature.

To carry out the biomechanical test, the installations of the mechanical test laboratory of the Nuclear Technology Development Center of the National Nuclear

Energy Commission (CDTN/CNEN) were used. The specimens were subjected to axial compression loads and flexion loads in the Instron TTDML® universal test machine (Canton, MA, USA), with a maximum capacity of 10 tons (Figure 2).

Metal supports were constructed to ensure a perfect fit for the specimens in the test machine while applying the loads, with the aim of avoiding any type of movement that could falsify the point of mechanical failure. In the compression test, no contact between the supports and the synthesis material was allowed, in order to avoid inappropriate transmission of the loads that were applied to the combined plate and bone. In the flexion test, it was sought to apply the load to the middle third of the diaphysis, on the cortex diametrically opposite the synthesis material (Figure 3).

The specimens were randomized into four groups according to the type of fixation and the type of test used. Thus, 20 specimens with fixation using blade plates and 15 using DCS were subjected to flexion loads, while 15 with fixation using blade plates and 15 using DCS were subjected to axial compression loads.

After correctly positioning the specimens, they were subjected to loads that were progressively increased at a rate of one centimeter per minute (cm/min). The loads, which were measured in kilogram-force (kgf), were plotted on analogue graphs as far as the point of mechanical failure. This was defined as a change in the

shape of the graph of force versus mechanical failure. The failures observed during the tests occurred either in the bone (fracture) or in the proximal screws of the plates (loosening), and were duly noted and statistically analyzed.

Statistical methodology

All the variables were subjected to descriptive analysis. Student's t test for independent samples was used to compare the mean forces to which the plates were subjected. The Anderson-Darling test was used to investigate whether the data presented normal distribution and the Levene test was used to investigate the supposition of equality of variance. Pearson's chi-square test was used to compare the proportions of bone failure and screw failure for each type of plate. The significance level was taken to be 5%.

RESULTS

The mean flexion force for the DCS plate was 89 kgf (SD = 22), with a minimum of 40 kgf and a maximum of 125 kgf. For the blade plate, the mean for this force was 109 kgf (SD = 41), with a minimum of 55 kgf and a maximum of 185 kgf (Table 1).

The boxplots (Figure 4) present the distributions of the flexion and compression forces according to the type of plate. The distribution of the flexion force was much more homogenous in specimens with DCS than

those with blade plates (smaller box, thereby indicating data of greater homogeneity in the sample). On the other hand, for the compression force, it was seen that the blade plates presented distribution of greater homogeneity than did the DCS.

The variables of flexion and compression forces as a function of type of plate did not violate the supposition of normality, according to the Anderson-Darling test. The supposition of equality of variances was not assumed, according to the Levene test.

Table 2 presents the test to compare the means of the flexion and compression forces as a function of the type of plate. No statistically significant difference in the mean flexion force (t = -1.904; p = 0.066) and mean compression force (t = 0.398; p = 0.694) according to type of plate was detected, i.e. the two plates were the same with regard to the flexion and compression forces.

To investigate whether the proportions of types of failure were different according to type of plate, Pearson's chi-square test was performed. In the tests using flexion force, it was observed that the assemblages with DCS failed visually more often in the bone, at the location of the osteotomy, while the assemblages with blade plates failed more often in the metadiaphysis region, at the location of screw fixation. This difference was not statistically significant (?2 = 2.37; p = 0.123), and it was concluded that the two plates behaved similarly with regard to types of failure (Table 3).

In the tests using compression force, it could be seen that the assemblages with DCS failed more often in the metadiaphysis region at the location of screw fixation, while the assemblages with blade plates failed more often in the bone, in the region of the osteotomy. This difference was not statistically significant (?2 = 0.52; p = 0.472) (Table 3).

DISCUSSION

Although the evolution of the treatment of supracondylar fractures of the femur follows the principles of rigid internal fixation, thereby enabling early rehabilitation(9), there is no consensus regarding the ideal fixation material(10-12).

In 1995, Firoozbakhsh et al(6) published a study comparing intramedullary rods with plates angled at 95º. They used synthetic femurs in which a medial defect had been produced (osteotomy), which were fixed using an intramedullary rod or a 95º degree blade plate. The results showed that the plate presented greater rigidity in relation to forces in the valgus direction, torsion and lateral flexion. On the other hand, the difference in compressive forces relating to the varus direction and medial flexion was not statistically significant.

In 1996, Koval et al(7) published a study that was similar to that of Firoozbakhsh et al(6). They concluded that the 95º plates presented greater rigidity than did the intramedullary rods, both in transverse fractures and in osteotomy of the distal femur with a medial defect.

In a study in 1997, David et al(5) compared intramedullary rods and DCS in several patterns of fracture of the distal femur. For this, they used femurs from human cadavers with different patterns of osteotomy. Through this, they produced different fractures that, based on the AO/ASIF group classification(8), ranged progressively in degree of severity from type 33ª to type 33C. The results showed that there was no difference between intramedullary rods and DCS according to the type of fracture and, consequently, they concluded that the choice of osteosynthesis material should not be based on the fracture pattern. They also recommended from this study that, if DCS were chosen, it should have 12 holes in a configuration with greater dispersion of the screws because this would have greater rigidity. On the other hand, if intramedullary rods were chosen, a configuration with closer grouping of the locking screws would be chosen.

One of the disadvantages of using DCS is the loss of bone material while preparing the hole for the dynamic screw(2). This finding was not confirmed by the present study, since no difference was observed between the two types of implant with regard to the type of failure, i.e. whether there was bone or implant failure (p = 0.123 in the flexion test and p = 0.472 in the compression test).

In a biomechanical test on DCS and 95º blade plates, Jaakkola et al(13) concluded that DCS presented greater rigidity in axial compression and a greater mean for the maximum load than observed with blade plates, which is in line with the results from the present study, although without statistical significance (p = 0.694). On the other hand, the results from the flexion test were similar, although in our study the blade plates presented a greater mean load than seen with DCS, albeit without statistical significance (p = 0.066). This trend might be proved statistically by increasing the sample size.

Harder et al(14) did not observe any difference in a biomechanical comparison between blade plates and DCS, with regard to treating unstable supracondylar fractures of the femur. However, these authors suggested that DCS should be used as the implant method of choice because of its technical ease, particularly if the distal fragment were at least 4 cm in size and in situations with less experienced surgeons. This result resembles the findings from the present study regarding the biomechanical equivalence of the plates, although no implant method that could be considered to be the one of choice was seen in the present study, despite the greater technical difficulty observed in introducing blade plates.

CONCLUSION

There was no statistically significant difference in relation to load resistance under flexion and compression, or in relation to the type of failure, i.e. whether it occurred in the bone (fracture) or in the material (loosening or breakage of the implant) between blade plates and DCS. However, there was an indication (p = 0.066) that blade plates might present greater rigidity in flexion than seen with DCS. We might be able to confirm this tendency if we were to increase the sample size of the blade plate group in the flexion test. In this study, we tested the rigidity of the plate-bone combination (measured in kgf) and not the fatigue of the osteosynthesis material. It needs to be emphasized that studies comparing these types of plate are scarce. Nonetheless, our findings are compatible with studies in the existing literature, in terms of equivalence between DCS and blade plates for treating extra-articular supracondylar fractures of the femur in experimental models.


REFERENCES

1. Rockwood CA, Green DP, Bucholz RW, Heckman JD, editors. Rockwood and Green's fractures in adults. 6th ed. Philadelphia: Lippincott Williams & Wilkins; 2006.
2. Schatzker J. Fractures of the distal femur revisited. Clin Orthop Relat Res. 1998;(347):43-56.
3. Albert MJ. Supracondylar fractures of the femur. J Am Acad Orthop Surg. 1997;5(3): 163-71.
4. Ito K, Grass R, Zwipp H. Internal fixation of supracondylar femoral fractures: comparative biomechanical performance of the 95-degree blade plate and two retrograde nails. J Orthop Trauma. 1998;12(4):259-66.
5. David SM, Harrow ME, Peindl RD, Frick SL, Kellan JF. Comparative biomechanical analysis of supracondylar femur fracture fixation: locked intramedulary nail versus 95-degree angled plate. J Orthop Trauma. 1997;11(5):344-50.
6. Firoozbakhsh K, Behzadi K, DeCoster TA, Moneim MS, Naraghi FF. Mechanics of retrograde nail versus plate fixation for supracondylar femur fractures. J Orthop Trauma. 1995;9(2):152-7.
7. Koval KJ, Kummer FJ, Bharam S, Chen D, Halder S. Distal femoral fixation: a laboratory comparison of the 95o plate, antegrade and retrograde inserted reamed intramedullary nails. J Orthop Trauma. 1996;10(6):378-82.
8. Müller ME, Allgower M, Schneider R. Manual of internal fixation. New York: Springer- Verlag; 1979.
9. Silisk J M, Mahring M, Hofer P. Supracondylar-intercondylar fractures of the femur - treatment by internal fixation. J Bone Joint Surg Am. 1989;71(1):95-104.
10. Arneson TJ, Meltron LJ 3rd, Lewallen DG, O'Fallon WM. Epidemiology of diaphyseal and distal femoral fractures in Rochester, Minnesota, 1965-1984. Clin Orthop Relat Res. 1988;(234):188-94.
11. Koval KJ, Hoehl JJ, Kummer FJ, Simon, JA. Distal femoral fixation: a biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate. J Orthop Trauma. 1997;11(7):521-524.
12. Sanders R, Swiontkowski M, Rosen H, Helfet D. Double-plating of comminuted, unstable fractures of the distal part of the femur. J Bone Joint Surg Am. 1991;73(3):341-6.
13. Jaakkola JI, Lundy DW, Moore T, Jones B, Ganey TM, Hutton WC. Supraconylar femur fracture fixation: mechanical comparison of the 95° condylar plate and screw versus 95° angled blade plate. Acta Orthop Scand. 2002;73(1):72-6.
14. Harder Y, Martinet O, Barraud G-E, Cordey J, Regazzoni P. The mechanics of internal fixation of fractures of the distal femur: a comparison of the condylar screw (DCS) with the condylar plate (CP). Injury. 1999;30(Suppl 1):A31-9..